E-line power on/off

Turn off

1. Login as administrator on pc (see [fig:E-line-computer])

2. Shut down gun.

3. Shut down EQ software.

4. Shut down EM-server program.

5. Turn off pc.

6. Press the yellow button on the side of the E-line (see [fig:E-line-tool]).

7. Take out airlock wire on the other side.

8. Press the red button besides the yellow (see [fig:E-line-tool]).

9. (Can turn of the main switch, big black one).

Turn on

1. (If the main switch is off, turn it on).

2. Plug in the airlock wire.

3. Press the yellow button - wait one minute.

4. Press green button after waiting one minute.

5. Wait for computer to boot up.

6. Log in as admin?