Temescal FC-2000 Electron-beam Evaporator: Difference between revisions

From Wiki for UiB's NanoStructures Laboratory
No edit summary
Line 3: Line 3:
===TC pump wont turn on===
===TC pump wont turn on===


If the TC pump is turned off it is mostlikly because the cryopump is to hot. Look at the stage two temperature on the system map. If it is above 20 Kelvin the TC can not be turned on. When this happens the system needs to be regenerated.
If the TC pump is turned off it is mostlikly because the cryopump is to hot. Look at the stage two temperature on the pump (insert image). If it is above 20 Kelvin the TC can not be turned on. When this happens the system needs to be regenerated.
 
====How to regenerate the system====
 
#Go to system map.
#Press detail in the upper right corner(insert image).
#When detail is pressed click on the cryopump (insert image).
#In the menu that pops up press regen SC, this should start the regenration prosses. It will take around 6 hours to finish.
#To see that it finished look at the crypump again, the stage two temperature should be around 11 kelvin and the TC should be on.


=== Anything else?===
=== Anything else?===

Revision as of 11:27, 18 October 2011

Normal problems with the Temescal

TC pump wont turn on

If the TC pump is turned off it is mostlikly because the cryopump is to hot. Look at the stage two temperature on the pump (insert image). If it is above 20 Kelvin the TC can not be turned on. When this happens the system needs to be regenerated.

How to regenerate the system

  1. Go to system map.
  2. Press detail in the upper right corner(insert image).
  3. When detail is pressed click on the cryopump (insert image).
  4. In the menu that pops up press regen SC, this should start the regenration prosses. It will take around 6 hours to finish.
  5. To see that it finished look at the crypump again, the stage two temperature should be around 11 kelvin and the TC should be on.