==Normal problems with the Temescal==
===TC pump wont turn on===
===TC pump wont turn on===


If the TC pump is turned off it is mostlikly because the cryopump is to hot. Look at the stage two temperature on the menue. If it is above 20 Kelvin the TC can not be turned on. When this happens the system needs to be regenerated.
If the TC pump is turned off it is mostlikly because the cryopump is to hot. Look at the stage two temperature on the menue. If it is above 20 Kelvin the TC can not be turned on. When this happens the system needs to be regenerated.

Revision as of 11:23, 18 October 2011

Normal problems with the Temescal

TC pump wont turn on

If the TC pump is turned off it is mostlikly because the cryopump is to hot. Look at the stage two temperature on the menue. If it is above 20 Kelvin the TC can not be turned on. When this happens the system needs to be regenerated.